UltiKleen™-CDS, Emflon® PF, Ultipleat P-Emflon® and Emflon® filter cartridges and assemblies require prewetting for the filtration of most acids and aqueous-based liquids in order to achieve good flow rates. Prewetting is not essential for solvents and liquids with surface tensions below approximately 30 dynes/cm2. Recommended procedures are as follows: UltiKleen™-CDS, Emflon® PF, Ultipleat P-Emflon and Emflon
NotePrewetting can be best accomplished by vacuum-drawing or pumping the 60/40 IPA/water prewetting solvent through the filter, followed by a 30-minute soak. This is especially recommended for the 0.05 µm filters, which are more difficult to wet due to tighter pores. If using concentrated acids with a high heat of mixing with water, it is strongly recommended that the filter be exposed to dilute acid (< 10%) prior to exposure with the concentrated acid. This prevents localized high temperature excursions from occurring. Top Emflon® PF Assembly
NoteEnsure that the assembly is vented correctly during the prewetting procedure. The presence of any trapped air in the assembly will cause the membrane to de-wet. V1 and V2 should be all fluoropolymer diaphragm valves. The pump, tubing and piping (upstream of the filter) should also be all fluoropolymer materials. Top |
||||||