| Product |
Acro® 37 TF Vent Devices with PTFE Membrane |
Bacterial Air Vents with Glass Laminate |
Acro® 50 Vent Devices with Emflon® II Membrane |
Acro® 50 Vent Devices with PTFE Membrane |
| Applications |
Small-volume venting and degassing. |
For venting and air/gas delivery. Exhibits higher temperature and pressure ratings, and higher air flow rates. |
Ideal for use in air/gas and vent applications with proprietary, low pressure drop hydrophobic PVDF membrane. |
Venting bioreactors, fermentation tanks, and carboys; sterile gas purging of culture vessels. |
| Materials of Construction Filter Media |
PTFE on a polypropylene support |
Hydrophobic glass laminate (polyester/glass fiber/polyester) |
Hydrophobic polyvinylidene fluoride (PVDF) |
PTFE on a polypropylene support |
| Housing |
Polypropylene |
Polypropylene |
Polypropylene |
Polypropylene |
| Pore Size |
0.2 µm |
1 µm (nominal) |
0.2 µm |
0.2 µm, 0.45 µm, 1 µm |
| Effective Filtration Area |
7.5 cm2 |
7.5 cm2 |
20 cm2 |
19.6 cm2 |
| Inlet/Outlet Connections |
Stepped hose barbs, 6.4 - 9.5 mm (1/4 - 3/8 in.) |
Stepped hose barbs, 6.4 - 9.5 mm (1/4 - 3/8 in.) |
Stepped hose barbs, 6.4 - 12.7 mm (1/4 - 1/2 in.) diameter with slip luer ID in the hose barb |
Stepped hose barbs, 6.4 - 12.7 mm (1/4 - 1/2 in.) diameter; 1/8 in. MNPT; 9.5 cm (3/8 in.) straight pipe |
| Max Operating Temperature |
100° C (212° F) at 1 bar (100 kPa, 15 psi) |
121° C (250° F) at 1 bar (100 kPa, 15 psi) |
60° C (140° F) at 1 bar (100 kPa, 15 psi) |
130° C (266° F) at 1 bar (100 kPa, 15 psi) |
| Max Operating Pressure |
4.1 bar (410 kPa, 60 psid) at ambient temperature |
5.2 bar (520 kPa, 75 psi) at ambient temperature |
4.1 bar (410 kPa, 60 psi) at ambient temperature |
4.1 bar (410 kPa, 60 psi) at ambient temperature |
| Typical Air Flow Rate |
3.58 L/min at 0.2 bar (20 kPa, 3 psi) |
40 L/min at 0.4 bar (40 kPa, 5.5 psi) |
27 L/min at 1 bar (100 kPa, 15 psi) |
L/min at 0.2 bar (20 kPa, 3 psi): 0.2 µm: 8 0.45 µm: 12 1 µm: 15 |
| Product |
AcroPak 300 Capsule with PTFE Membrane |
HEPA Capsule |
Vacushield® Vent Device |
|
| Applications |
Bioreactor venting requiring high air flow rates. |
Provides bacteria-free air for sterile applications. |
Use between a pump and receiving vessel to protect valves and pump components from damage from aqueous solutions; also protects lab personnel from aerosolized contaminants. |
| Materials of Construction Filter Media |
PTFE on a polypropylene support |
Versapor® membrane (acrylic copolymer on a nonwoven support) |
PTFE on a polypropylene support |
| Housing |
Polypropylene |
Polypropylene |
Polypropylene |
| Pore Size |
0.2 µm |
1.2 µm (liquid filtration); 99.97% retention of 0.3 um DOP aerosol (following ASTM D 2986-95A) |
0.45 µm |
| Effective Filtration Area |
300 cm2 |
860 cm2 |
19.6 cm2 |
| Inlet/Outlet Connections |
Stepped hose barbs, 6.4 - 12.7 mm (1/4 - 1/2 in.) diameter |
3/8 in. FNPT, includes optional hose barb fittings to accept 12.7 mm (1/2 in.) ID tubing |
Stepped hose barbs 6.4 - 12.7 mm (1/4 - 1/2 in.)diameter; internal taper accepts standard male luer |
| Max Operating Temperature |
60° C (140° F) at 2.1 bar (210 kPa, 30 psi) |
88° C (190° F) |
130° C (266° F) |
| Max Operating Pressure |
4.1 bar (410 kPa, 60 psi) at ambient temperature |
3.4 bar (340 kPa, 50 psi) |
1 bar (100 kPa, 15 psi) |
| Typical Air Flow Rate |
32 L/min at 0.07 bar (7 kPa, 1 psi) |
|
8 L/min at 0.1 bar (10 kPa, 2 psi) |