Pall Corporation Homepage: Filtration, Separations, Purification, and Environmental Technology
Check Out Lab Store
My Account Quick Order
Home
1.800.521.1520 (USA)
EMAIL USLIVE HELPCONTACT US

Cascada™ BIO-Water Systems: Compliance With BS3978/ISO 3696 Specification for Water for Laboratory Use, Grade 1

Introduction

The Cascada BIO-water System is fed with partially purified water. It produces extremely high ultrapure water for life science applications, suitable for all molecular, pharmaceutical, and microbiological applications.

The product water conforms to BS3978/ISO 3696 Grade 1 specification.


Top

Results

Parameter Grade 1 Specification Typical Measured Values
Resistivity (Mohm-cm at 25 °C) > 10 18.2
Silica (ppb) < 500 < 1
Absorbance at 254 nm < 0.001/cm < 0.001/cm

Top

Method

A Cascada BIO-water System was fed with water from a Cascada RO-water System with a conductivity of 42 µS/cm and a TOC of 50 ppb C. The unit was operated intermittently and product water dispensed to simulate typical use patterns.

Resistivity was logged online using a Thornton 200CR meter. Samples were taken for silica analysis and absorbance measurements at various times.

Samples were analyzed for silica by ICP-MS using a PlasmaQuad 3.

View additional information on water purification systems and accessories.


Top